High-level disinfection of endoscopes and instruments. 5-minute exposure time.
A two-component high-level disinfectant based on peracetic acid for flexible endoscopes and medical instruments. Mixing the base component with the activator produces a working solution with a peracetic acid concentration of 2200 ppm. Provides bactericidal, fungicidal, virucidal, tuberculocidal and sporicidal action in just 5 minutes. Free of aldehydes and phenols; suitable for glass, silicone, plastic, metal and rubber.
Pour the activator into the base bottle, close and shake well — the solution is ready in 10 minutes; pour the solution into a clean container and fully immerse pre-cleaned instruments, filling all cavities; close the container and maintain the exposure time; rinse the instruments with sterile or distilled water; check the concentration daily with test strips (do not use if the concentration falls below 900 ppm peracetic acid); the solution is valid for 28 days. Do not use on aluminium, nickel or chrome-plated surfaces.