High-level disinfection of endoscopes: the steps
Endoscopes contact mucous membranes and cannot be processed with high heat. High-level disinfection (HLD) with chemical agents in a strict sequence is therefore used.
Reprocessing steps
- Pre-clean immediately after the procedure to remove organic soil;
- Leak test;
- Manual or automated cleaning with an enzymatic agent, including channels;
- High-level disinfection with an agent of confirmed activity;
- Rinse with clean (ideally sterile) water;
- Dry the channels and outer surfaces;
- Store in conditions that prevent recontamination.
Why cleaning before HLD matters
Biofilm and organic residues shield microorganisms from the disinfectant. Without proper pre-cleaning, even an effective agent will not work — cleaning and HLD work only together.
Quality control
Correct concentration, contact time and solution temperature are critical. Keeping reprocessing logs and using control tools makes the process reproducible.